UV/Vis Spectrophotometers

UVI LIGHT Specifications
Visible light source Pre-adjusted Halogen lamp
UV light source Pre-adjusted Deuterium lamp
Monochromator Jobin Yvon concave holographic grating, resolution 1200 lines/mm
Detector Silicon diode
Method storage Local storage for 50 methods
Wavelength range 190 - 900 nm
Wavelength increment 0.1 nm 
Spectral bandwidth 2 nm fixed.
Wavelength accuracy ± 1 nm
Wavelength precision
Holmium Oxide Filter
± 0.1 nm
Baseline stability < ± 1 mabs
Stray light < 0.4% at 200 nm, < 0.002% at 220 and 340 nm
Photometric range -0.1 to 3 AU, 0.1 to 110 %T
Photometric accuracy ± 5 mabs
Drift < 3 mAbs/hr measured at 500 nm after warm-up
Dimensions 500 x 295 x 415 mm (W x H x D)
Weight 7.5 kg