Wavelength range
|
190 - 1100 nm
|
|
Wavelength step
|
0.05 - 10 nm
|
|
Scan speed
|
5 - 2000 nm/min
|
|
Transfer speed
|
5000 nm/min
|
|
Spectral bandwidth
|
1.8 fixed
|
Wavelength accuracy
Holmium Oxide filter
10nm/min, 0.05nm step
|
± 0.3 nm
|
Wavelength precision
Holmium Oxide Filter. Shoulder at 444nm, 2nd wavelength 536.7nm. Speed 5000nm/min
|
± 0.03 nm
|
|
Baseline flatness. 200-800nm, 200nm/min, 1nm step
|
<± 1 mAbs
|
Stray light
Nal (10g/i), 220 nm, 1sec
|
< 0.03 %T
|
Linearity
250 nm, 1 sec.
|
3.0 Abs, (r
2
>= 0.999)
|
|
Photometric range
|
± 3.5 Abs
|
|
Photometric accuracy. 1.0Abs certified filter,1sec, 590nm
|
± 3 mAbs
|
Photometric precision.
1.0 Abs certified filter, 1sec, 590nm.
|
± 0.5 mAbs
|
|
Noise (RMS), measured at 580nm, 1sec, ASTM E 685
|
better than 30 µ Abs
|
|
Drift, 580nm, 1 sec after 2 hours warm up
|
< 0. 1 mAbs/hr
|
|
Response time
|
0.02 - 5 sec
|
|
Light sources
|
Deuterium and Tungsten-Halogen lamps
|
|
Monochromator
|
High energy low stray light diffraction grating with 1300 lines/mm, 175 mm focal length
|
|
Dimensions
|
680 x 275 x 565 mm (W x H x D)
|
|
Weight
|
35 kg
|
|