Wavelength range
|
180 - 900 nm
|
Wavelength step
|
0.05 - 10 nm
|
Scan speed
|
5 - 5000 nm/min
|
Transfer speed
|
7000 nm/min
|
Spectral bandwidth
|
0.2,0.5,1,2,4nm and 0.5,1,2,4nm reduced height
|
Wavelength accuracy
Holmium Oxide filter
10nm/min, 1nm, 0.05nm step
|
± 0.25 nm
|
Wavelength precision
Holmium Oxide Filter. Shoulder at 444nm, 2nd wavelength 536.7nm. Speed 7000nm/min
|
± 0.025 nm
|
Baseline flatness. 200-800nm, 2nm, 200nm/min
|
± 1 mAbs
|
Stray light
Nal, 220 nm, 1nm, 1sec
|
< 0.015 %T
|
Linearity
250 nm, 2 nm, 1 sec.
|
3.3 Abs, (r
2
>= 0.999)
|
Photometric range
|
± 5 Abs
|
Photometric accuracy. 0.85 Abs certified filter,1sec, 590nm
|
± 3 mAbs
|
Photometric precision. 0.85 Abs certified filter, 1sec, 590nm.
|
± 0.5 mAbs
|
Noise (RMS), measured at 580nm for 5mins, 2nm, 1sec
|
40 µ Abs
|
Drift, 580nm, 2nm, 1 sec after 2 hours warm up
|
< 0. 1 mAbs/hr
|
Response time
|
0.02 - 10 sec
|
Light sources
|
Deuterium and Tungsten-Halogen lamps
|
Monochromator
|
High energy low stray light diffraction grating with 1300 lines/mm, 175 mm focal length
|
Dimensions
|
680 x 275 x 565 mm (W x H x D)
|
Weight
|
35 kg
|